From: Method development and validation of potent pyrimidine derivative by UV-VIS spectrophotometer
Serial number | Parameters | Acceptance criteria | Observation |
---|---|---|---|
01 | Precision | ||
(a) System precision %RSD | NMT 1.5% | 0.0968 | |
(b) Method precision %RSD | NMT 1.5% | 0.27995 | |
02 | Specificity | No considerable absorbance of any other component of formulation at λmax of analyte or at detection wavelength | No absorbance observed at 275 nm |
03 | Accuracy (by recovery) | ||
% Recovery | 100% ± 2% | 100.12% | |
%RSD | NMT 1.5% | 1.1777% | |
% Deviation from accuracy | ±1.5% | 080%: +01.46 | |
100%: −00.77 | |||
120%: −00.32 | |||
04 | Linearity | Coefficient of correlation | r2: 0.997 |
(r2) NLT 0.998 | |||
05 | Ruggedness | %RSD: NMT 1.5% | %RSD: 0.1572 |
06 | Robustness | %RSD: NMT 1.5% | Original condition: +0.72 |
Changed condition: −0.82 | |||
07 | Limit of detection and limit of quantitation | LOD | 145.2 mg |
LOQ | 440.00 mg |